CVE-2019-25720

MEDIUM

CVSS v3.1: 6.5

No privilegesNo user interactionVendor advisory ref

Source data as of:

At a glance

Severity
MEDIUM
CVSS
6.5 v3.1 · NVD
EPSS
EPSS not provided by FIRST.org for this CVE
CISA KEV
No
Attack conditions (CVSS vector)
No privilegesNo user interaction · Source: NVD Vector
Published
2026-06-03 · Modified: 2026-06-03

CVSS / EPSS / KEV

CVSS v3.1 6.5 / 10 MEDIUM Source: NVD
CVSS v4.0 7.1 / 10 HIGH Source: NVD
EPSS EPSS not provided by FIRST.org for this CVE Source: FIRST.org
CISA KEV No Source: CISA

Source — CVSS: NVD · EPSS: FIRST.org · KEV: CISA. Data & Sources

Description

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

Record details

CVE ID
CVE-2019-25720
CVSS (v3.1)
6.5 (MEDIUM)
Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Exploitability subscore
2.8
Impact subscore
3.6
CISA KEV
No
Weakness (CWE)
CWE-1286
Affected configurations (CPE)
0
Published
2026-06-03
Modified
2026-06-03
Status
Received

References

Reference URLs as listed by NVD, grouped by a mechanical match on the link's host/pattern. Labels describe the link type only.